The potency of vaccines degrades quickly when they aren’t stored within the correct temperature range. Monitoring and recording of temperatures within refrigerators, freezers and containers used for storing vaccines is therefore well-established practice. However, ensuring that readings taken match the actual temperatures experienced by vaccines requires the methodical application of best practice. Failure to do so can lead to the administering of vaccines that lack the necessary potency to ensure effective immunization, or the unnecessary disposal of valuable stock.
The Vaccine Cold Chain
A cold chain is a temperature-controlled supply chain that includes all vaccine-related equipment and procedures. The cold chain begins with the cold storage unit at the manufacturing plant, extends to the transport and delivery of the vaccine and correct storage at the provider facility, and ends with administration of the vaccine to the patient.
– Centers for Disease Control and Prevention (CDC)
Vaccines are active biological products that are damaged to varying extents by exposure to high and low temperatures. Once vaccine potency is lost, it cannot be regained. This may result in the failure of the vaccine to create the desired immune response and consequently it will provide poor protection. Unfortunately, vaccines do not change appearance when they have been compromised by temperatures outside the acceptable range. Only the continual use of suitable storage equipment and procedures can guarantee vaccine effectiveness.
Improving Temperature Stability
Within the refrigerator, sufficient space around the vaccine packages should be left for air to circulate. Vaccines should be kept away from the side and back walls of the refrigerator; otherwise the vaccines may freeze rendering them inactive and unusable.
– Public Health England, Green Book
Even when using refrigeration equipment specifically designed for medical use, vaccines should be stored so as to maximize temperature stability. Common points in accepted best practice worldwide include:
- Store each type of vaccine in its original packaging and in a separate container.
- Position vaccines two to three inches from the unit walls, ceiling, floor, and door.
- Avoid storing vaccines in any part of the unit that may not provide stable temperatures or sufficient air flow, such as directly under cooling vents, in drawers, or on refrigerator door shelves.
- Avoid placing or storing any items other than vaccines, diluents, and water bottles inside storage units.
- If other medications and biological products must be stored in the same unit as vaccines, they must be clearly marked and stored in separate containers or bins from vaccines.
- Arrange vaccines in rows and allow space between them to promote air circulation.
For some refrigerators, placing water bottles in the unit can help maintain stable temperatures despite frequent opening and closing of unit doors, or in a power failure. Where used, these bottles should be placed on the top shelf and floor, and in the door racks.
Temperature Monitoring Equipment
CDC recommends a specific type of temperature monitoring device called a “digital data logger” (DDL). A DDL provides the most accurate storage unit temperature information, including details on how long a unit has been operating outside the recommended temperature range (referred to as a “temperature excursion”).
– Centers for Disease Control and Prevention (CDC)
Although refrigerators and freezers will contain their own temperature regulator, which may also record data, it is not acceptable practice to rely solely on this as it constitutes a single point of failure. The digital data logger used must be independent of the mains power supply to the fridge itself and so will be powered by batteries. Key features for accurate and reliable temperature monitoring of stored vaccines include:
- Accurate within ± 0.5°C (± 1°F)
- Display showing current, minimum, and maximum temperatures
- Alarm for out-of-range temperatures
- Logging interval, programmable by the user, to measure and record temperatures at least every 30 minutes
- Low-battery indicator
- Detachable buffered probe that accurately reflects vaccine temperatures
Potentially damaging fluctuations in temperature can be hidden if recording intervals are longer than a few minutes, and so effective data loggers will combine short logging intervals with a memory capable of storing long time periods logged at such intervals.
All data loggers should also be supplied with a calibration test certificate to national standards.
Buffered Temperature Probes
A temperature probe designed to prevent false readings by protecting the thermometer from sudden changes in temperature that can occur when opening a refrigerator door.
– Centers for Disease Control and Prevention (CDC)
A critical but often overlooked aspect of vaccine temperature monitoring is that it’s not enough simply to measure the temperature of the air inside the fridge or freezer. Air gets disturbed or replaced each time the door is opened, and also has a very different thermal inertia to vaccines themselves. Glycol has similar thermal properties to vaccine preparations and so using a probe sealed into a bottle of glycol ensures the readings taken accurately reflect the temperatures experienced by the vaccines and not just the air in the fridge.
The following diagram illustrates the difference between unbuffered and buffered probes in a medical fridge with the usual 8°C upper temperature limit for vaccine storage.
Comparison between buffered and unbuffered temperature probes
A: Fridge door briefly opened – vaccine not exposed to high temperatures
B: Fridge door opened – unbuffered probe triggers false alarm
C: Fridge door left open – buffered probe correctly triggers alarm when vaccine exceeds safe storage temperature, unbuffered probe triggers early false alarm and misses critical time at which safe temperature is exceeded
Calibration and Maintenance
WHO recommends that temperature monitoring devices and control sensors that are designed for recalibration be calibrated against a certified, traceable reference standard at least once a year unless otherwise justified.
– World Health Organization (WHO)
Calibration is a test process that confirms the accuracy of an individual data logger, distinct from the designed accuracy of that model of logger. Temperature loggers should therefore be supplied from new with a calibration test certificate that references the serial number of that unit. Due to the inevitable drift in the performance of individual components within the logger, calibration testing should be repeated at the frequency specified by the manufacturer, typically every 12 months. Modern digital data loggers should require no other maintenance, other than the replacement of batteries as soon as the low battery indicator activates. To ensure continuity in temperature records, a supply of spare batteries should be kept on hand at each location loggers are used.
Conclusions
There are two key modes of failure in vaccine cold chain storage, and both can be prevented by keeping accurate temperature records.
The first is that vaccines exposed to too high or too low temperatures, for a long enough period to degrade potency, can be identified and removed from use.
The second is to prevent the unnecessary disposal of vaccine stocks that appear to have a suffered loss of potency, but are actually undamaged and will still provide effective immunization.
The costs associated with the loss and replacement of compromised vaccines should not be underestimated. According to official figures, during 2018 vaccine wastage in the UK alone had a value of at least £6.3 million. Public Health England accepts this is likely to be underreported and thus the true financial cost even greater.
References:
World Health Organization Vaccine Management Handbook (WHO/IVB/15.04)
Immunization Against Infectious Disease (‘The Green Book’), Public Health England, revised 2nd September 2014
CDC Vaccine Storage and Handling Toolkit, US Department of Health and Human Services, January 2019
Vaccine Incidence Guidance, Public Health England, September 2019